FI
(top)
< 0.15 pM FITC (384sv black, 20 μL)
< 0.5 pM FITC (1536 black, 8 μL)
Dynamic Range: 8 decades
FI
(bottom)
<1.0 pM FITC (384g black, 50 μL)
FP
< 0.5 mP SD at 1 nM FITC (384sv black, 20 μL)
< 1.5 mP SD at 1 nM FITC (1536 black, 8 μL)
TRF with laser.
< 5 fM europium (384, 80 μL)
< 15 fM europium (1536, 8 μL)
HTRF®
(black and white microplates)
Reader Control Kit (Eu) after 18h (384sv, 20 μL)
Delta F > 1100% (High Calibrator)
Delta F > 25% (Low Calibrator)
LUM
< 0.4 pM (< 8 amol/well ATP, 384sv white, 20 μL)
Dynamic Range: 8 decades
AlphaScreen® with laser
< 5 pM (< 100 amol/well P-Tyr-100, 384sv white, 20 μL)
ABS with spectrometer
Selectable spectral resolution: 1, 2, 5, and 10 nm
OD range: 0 to 4 OD; photometric resolution: 0.001 OD
Accuracy: < 1% at 2 OD
Precision: < 0.5% at 1 OD and < 0.8% at 2 OD
Linearity: ≤ 0.8% at 2.0 OD