5 BMG LABTECH filters 5 BMG LABTECH filters

Fluorescence intensity (incl. FRET)

Measure fluorescence intensity with high sensitivity, enabling diverse applications from gene reporters to enzyme kinetics.

What is fluorescence intensity?

Fluorescence intensity is the amount of light emitted by a fluorescent substance after it absorbs light at a specific excitation wavelength. The emitted light is produced by fluorescent molecules called fluorophores (Fig. 1).

Fig. 1: Modified Jablonski diagram. In fluorescence, an energy source lifts (excitation) molecules into electronically excited states (S2). Falling back into ground state (S0), light (photons) is emitted, producing a fluorescent signal.

When a fluorophore absorbs light, its electrons move to a higher energy state. As the electrons return to their ground state, they release some of the absorbed energy as light. The emitted light usually has a longer wavelength than the excitation light. This difference is known as the Stokes shift (fig. 2).

Fig. 2: Stokes shift is the difference in wavelength between the excitation (absorption) and emission peaks.

 

In a suitable assay, fluorescence intensity can be related to the concentration of the fluorophore or analyte. This makes fluorescence measurements useful for quantitative analysis in applications such as cell viability testing, enzyme activity assays, nucleic acid quantification, protein studies, and drug discovery.

Fluorescence intensity measurements are commonly performed using microplate readers, fluorescence microscopes, and flow cytometers. This article focuses on fluorescence detection in microplate-based assays, including instrument configuration, measurement considerations, and common fluorescence assays.

 

Fluorophores and fluorescence properties

Fluorophores are molecules that absorb light at one wavelength and emit light at another. Their optical properties determine how they perform in a fluorescent assay and influence the sensitivity, selectivity, and compatibility of a fluorescence detection method.

When selecting a fluorophore, consider:

  • Molecular size and compatibility with the biological target

  • Excitation and emission wavelengths

  • Quantum yield and brightness

  • Photostability

  • Compatibility with the sample matrix and detection instrument

  • Spectral overlap with other fluorophores or sources of autofluorescence

Molecular size

Many fluorophores are small chemical compounds that can be linked to proteins, nucleic acids, antibodies, or other molecules. Fluorescent proteins, such as green fluorescent protein (GFP) and red fluorescent protein (RFP), are larger fluorescent tags that can be expressed by living cells.

Fluorescent proteins can be used to monitor protein expression, study cellular processes, develop genetically encoded biosensors, or estimate cell number. For example, GFP and mCherry can be used to monitor transfection efficiency.

Excitation and emission wavelengths

A fluorophore’s excitation and emission properties are shown using fluorescence spectra. Wavelength is plotted on the x-axis, usually in nanometres, while relative fluorescence intensity is plotted on the y-axis.

The spectra show the wavelengths at which the fluorophore absorbs and emits the most light. These maxima help determine the appropriate settings for a fluorescence assay.

When selecting a fluorophore, make sure its excitation and emission wavelengths are compatible with:

  • The filters or monochromators in the plate reader

     

  • The detector’s wavelength range

     

  • Potential autofluorescence from the sample or buffer

     

  • Other fluorophores used in multiplex fluorescence assays

Quantum yield

The quantum yield is a measure of how efficiently an excited molecule converts absorbed photons into emission light. In combination with the extent to which light is absorbed (extinction coefficient), the quantum yield gives an indication of the brightness of a fluorophore.

 

FRET pairs

Förster resonance energy transfer (FRET) describes the transfer of energy from a donor fluorophore to an acceptor fluorophore. The donor and acceptor must be sufficiently close for energy transfer to occur.

FRET assays are used to study molecular binding, conformational changes, receptor-ligand interactions, and biosensor responses. For FRET to occur, the donor emission spectrum must overlap with the acceptor excitation spectrum.

In a FRET biosensor, binding of an analyte can change the distance or orientation between the donor and acceptor. This produces a measurable change in fluorescence intensity or in the ratio between donor and acceptor fluorescence.3

Fig. 3: Example of the FRET fluorophore pair Cyan Fluorescent Protein (CFP) and Yellow Fluorescent Protein (YFP). The CFP (donor) emission spectrum overlaps with the YFP (acceptor) excitation spectrum (spectral overlap). 

 

How fluorescence detection works

Even though fluorophores differ from each other, the basis of fluorescence intensity measurements in microplates is the same (fig. 4):

  • Broadband light is produced by a light source
  • The excitation wavelength range is filtered from the broadband light
  • The excitation wavelength range excites the fluorophore
  • The emission wavelength range is filtered from the emission light
  • A detector quantifies the filtered emission light

As opposed to absorbance, fluorescence is not an absolute measurement. The intensity of the fluorescent signal is usually relative to other measurements or to a reference measurement taken by an instrument. Consequently, fluorescence plate readers measure the light signal emitted by a sample in Relative Fluorescent Units (RFU).

Fig. 4: simplified construction of a fluorescence intensity detection system in a microplate reader.

 

Light sources

The light source affects the available excitation wavelengths, signal strength, measurement speed, and instrument lifetime. Common light sources in microplate readers include xenon flash lamps, tungsten-halogen lamps, and light-emitting diodes. Read more in the HowTo Note: How does the number of flashes influence measurement results?

 

Xenon flash lamp

Xenon flash lamps produce a broad spectrum of light in the range of 200 nm to 2500 nm and can support fluorescence detection in the UV range to the infrared range. They exhibit a high signal output and are long-lived. Because of these advantages, xenon flash lamps are used as fluorescence excitation sources in BMG LABTECH plate readers.

 

Tungsten halogen lamp

Halogen lamps emit light starting from approximately 360 nm and are unsuited to measure UV fluorescence intensity. In comparison xenon lamps, tungsten lamps have a shorter lifetime, a lower output intensity, and result in lower sensitivity, if used in microplate reader fluorescence detection.

 

LEDs 

Light Emitting Diodes (LEDs) emit light with high intensity but only at a specific bandwidth. Accordingly, multiple diodes are required for different excitation wavelengths, making flexible detection of multiple fluorophores more complicated and expensive.

 

Wavelength selection

Selecting suitable excitation and emission wavelengths is essential for sensitive fluorescence detection. The excitation wavelength should efficiently excite the fluorophore, while the emission range should capture the emitted light and minimise the amount of excitation light reaching the detector. Fore more information read the Howto Note: Wavelength and bandwidth optimisation on a microplate reader.

The sensitivity of detection in some fluorescence intensity microplate readers is enhanced by the use of an additional selection between excitation and emission: a dichroic mirror. The dichroic mirror typically transmits light higher than a specific wavelength but blocks light at lower wavelengths. The cut-off lies between excitation and emission and provides lower leak-through of excitation light to the detector.

 

Optical Filters

Optical filters transmit a defined range of wavelengths and block most of the remaining light. A filter is often described by its centre wavelength and bandwidth.

For example, a 485 nm filter with a 10 nm bandwidth may transmit wavelengths from approximately 480 to 490 nm. The exact transmission range depends on the filter specification.

The excitation and emission filters should be selected using the fluorophore’s spectra and the requirements of the fluorescent assay. Filter choice should also account for autofluorescence, spectral overlap, and the sensitivity of the detector.

Fig. 5: different types of optic filters.

 

 

Monochromators

Unlike fixed optical filters, monochromators can select different wavelengths according to the measurement requirements. Microplate readers may use grating-based monochromators or Linear Variable Filter (LVF) monochromators.

Grating-based monochromators separate broadband light by diffraction. A slit then selects the required wavelength range. However, light transmission can be lower than with some filter-based systems, and stray light may reduce fluorescence sensitivity.

Linear Variable Filter (LVF) Monochromator use filters with a continuously varying transmission profile. By positioning the filters appropriately, the instrument can select different wavelengths and bandwidths without dispersing the light into separate colours. The use of Linear Variable Filters in microplate readers is a patented technology by BMG LABTECH.

Fig. 6: LVF Monochromator principle

 

Detectors for fluorescence measurements

Fluorescence intensity in microplate readers is detected with photomultiplier tubes (PMTs). They multiply the signal using the photoelectric effect and convert the light into an electric signal. For fluorescence intensity measurements in microplate readers, two different PMTs are used, which differ in their sensitive wavelength range. A low-noise blue/green PMT is sensitive in the range up to 740 nm and is often also used for luminescence measurements. A red-shifted PMT measures the fluorescence sensitively up to 900 nm and thus also covers red and infrared dyes. These are often used for cell-based measurements since the autofluorescence is reduced in this wavelength range.


How to set up a fluorescence measurement on a plate reader

The performance of a fluorescence assay depends on both the assay chemistry and the measurement settings. Plate type, optical configuration, gain, focal height, wavelength selection, incubation conditions, and measurement timing can all affect the fluorescence signal.

 

Microplate

Black microplates are generally preferred for fluorescence assays because they reduce reflection and the spread of excitation and emission light between wells. More details on the plate choice are found in our blog post "The microplate: utility in practice."

 

Filter and monochromator settings

Choosing the right filter is essential for a precise fluorescence intensity determination. As already mentioned, the mean wavelengths should be at or close to the excitation and emission maximum. However, there are a few things to keep in mind:

 

a) Distance between filters 

If the edges of excitation and emission filters are too close to each other, the excitation light may leak through and reach the detector. Then the signal of the fluorophore disappears in the excitation light and is no longer measurable. How close the filters can be, depends on the quality of the filters and on the use of a dichroic mirror. Typically, the highest transmitted excitation and the lowest transmitted emission wavelengths should be 30 nm apart (fig. 7).

Fig. 7: schematic of excitation and emission filters for fluorescein superimposed to their spectra. The highest transmitted excitation and the lowest transmitted emission wavelengths should be 30 nm apart.

 

b) Bandwidth

The bandpass of filters varies from a few nanometers up to 100 nm. The bandpass affects the amount of light that can pass the filter. Low bandwidth filters (~10nm) are recommended for bright fluorophores or for complex samples with high autofluorescence. A typical example is green fluorophores such as AlexaFluor488, FITC, or GFP in a cellular environment. 

Broader bandwidth filters are recommended for low emission fluorophores if they emit in a range where autofluorescence is low (emission <600nm) or if the buffer contains no other fluorescent compound. BMG LABTECH selects the filters based on the fluorophore excitation and emission spectra. In addition, filter pairs are measured and matched to one another so that you can be sure that you are receiving the best possible filters. Default settings for common fluorophores are also already stored for the monochromator so that this selection work is not left to the researcher.

 

c) FRET measurements require two emission channels

The above considerations also apply to FRET measurements. However, it must be noted that two emissions must be recorded separately. The excited donor fluorophore emits light at a specific wavelength if no acceptor is nearby. If an acceptor fluorophore is close, the donor transfers the energy to the acceptor which then emits light at a higher wavelength. This means that two emission filters are required, one of which only allows the donor emission and the other only the acceptor emission to be transmitted.

 

Gain and dynamic range

The fluorescence gain is an amplification factor, or more precisely, the voltage of the detector at which the incoming light signal is amplified. Changing the gain moves the dynamic range of detection along the concentration curve of the analyte/assay. However, the width of the dynamic range is fixed and does not change. 

Weak signals need a large amplification (high gain), and bright signals need less amplification (small gain). An incorrectly selected gain can make measured data unusable because it either exceeds the measuring range or differences are no longer resolved. Data highlighting these effects can be found in the HowTo Note: How to optimise the gain setting of your microplate reader? The gain is typically set to have a maximum measurement output on the sample with the expected highest intensity. This is done in order to have the largest possible dynamic window between the highest and the lowest measurement values (fig. 8). 

All BMG LABTECH microplate readers determine the gain automatically, provided a specific well with a specific signal intensity is identified. However, on the CLARIOstar Plus and VANTAstar, no gain adjustment is necessary. The Enhanced Dynamic Range (EDR) technology was specifically designed to offer the largest possible dynamic range - 8 concentration decades. This provides researchers with an unprecedented convenience in microplate measurements as highly reliable results can be measured over a large dynamic range with no manual intervention.Fig. 8: high gain values increase sensitivity, as they allow discerning between low signals and blanks, but can easily result in overflow in case a brighter signal is present. High signals require low gain values to avoid saturation of the detector, but limit sensitivity as the detector is no longer able to separate low signals from blanks.

 

Focal height

The height at which a signal is detected has a significant impact on the sensitivity of the measurement. Measuring at a suboptimal focal height may reduce the signal by up to 90 %. For instance, adherent fluorescent cells need to be measured close to the plate bottom, whereas a homogeneous fluorescent solution gives the highest signal just beneath the liquid surface and depends on the fill volume of the well. Some plate readers determine the optimal focal height in a selected well. The CLARIOstar® Plus and VANTAstar® also provide automatic focal-height optimisation and determine the focal height automatically while reading a plate.

 

Applications of fluorescence assays

Fluorescence assays use changes in fluorescence intensity to measure biological activity, molecular concentration, cellular responses, or interactions between biomolecules. The following examples show how fluorescence analysis is used in research and drug discovery.

 

Cell viability and cytotoxicity assays

Whether cells are harmed by a compound or cancer cells die upon treatment are common questions answered by fluorescent viability and cytotoxicity assays and microplate readers. In the alamarBlue™ assay, resazurin is reduced by metabolically active cells to a fluorophore and reports on cell viability. Another way to assess cell viability is the incorporation of fluorescently marked nucleosides into the DNA. These are incorporated and consequently fluoresce only if DNA replication takes place, a marker of viability. 

A way to detect cell death by fluorescence intensity is based on fluorescent DNA dyes such as SYBR® Green or Celltox™ Green. The dyes are non-cell-permeable but get exposed to DNA in solution in case the cell membrane is disrupted during cell death and then starts to fluoresce (fig. 9). These assays are compatible with real-time measurements and can monitor cell health over time. As these measurements are done over days, the microplate chamber needs to be incubated at 37 °C and with 5 % CO2 to keep cells happy. The CLARIOstar Plus with Atmospheric Control Unit has proven to provide the ideal atmosphere for long-term cell-based experiments. More cell health assays are described in our blog post “Cell viability assays – Measure how happy your cells are”.Fig. 9: the CellTox™ Green Cytotoxicity Assay Principle.

 

Enzyme activity assays

As enzymes are involved in multiple biological processes, it is important to study how inhibitors or promoters influence enzyme activities. The principles of enzyme assays are comparable: they employ a pre-fluorescent substrate which upon enzyme processing releases a fluorophore. Common fluorophores are based on coumarin derivatives such as in the fluorescence-based assay of the epigenetic enzyme histone deacetylase 1 (HDAC1), where they can be used to investigate inhibition of histone deacetylases.  

A second fluorophore used in enzyme assays is 4-methylumbelliferone. An example is found in the application note “CLARIOstar determines activity of a moss-produced human acid alpha-glucosidase (GAA) in a fluorescence-based assay”. The CLARIOstar Plus multi-mode microplate reader makes fluorescent enzyme activity analysis easy. The Enhanced Dynamic Range technology allows reading over a dynamic range of 8 decades. This way, the increasing signal in enzyme activity assays is easily recorded without the risk of choosing a wrong gain or time-consuming instrument optimization. The instrument comes with a free-of-charge analysis software with integrated enzyme kinetic analysis covering Michaelis-Menten and Lineweaver-Burk modelling (Enzyme kinetic measurements performed on a BMG LABTECH microplate reader).

 

Conformational changes: protein folding and unfolding

The fluorescence of naturally occurring amino acids like tryptophan can also be used to measure conformational changes and enzyme activity of proteins. Some examples of applications are described in our blog post Tryptophan fluorescence: nature’s probe. Tryptophan fluorescence is useful for example to screen for new therapeutics for drug targets like G-protein-coupled receptors, one of the most popular targets for approved drugs on the market today.

 

Protein aggregation assays

Proteins aggregate if they are misfolded and aggregation correlates with diseases such as Alzheimer’s, Parkinson’s, and prion disease. The analysis of the formation of molecular aggregates helps diagnose and understand these diseases, as well as test potential drugs. The two fluorophores bis-ANS and Thioflavin T (ThT) increase their fluorescence intensity upon interaction with protein aggregates. Bis-ANS preferably interacts with amorphous aggregates as described in the application note “Use of a BMG LABTECH microplate reader to monitor amyloid formation”. 

ThT binds to fibrils and is mainly used in kinetic assays. If the sample contains aggregation-inducing molecules, the recombinant proteins aggregate, and fluorescence increases (fig. 10).

Fig. 10: fibrillization process followed over time.

The application note “Real-time quaking-induced conversion assay for prion seeding” explains how the assay is used for detecting the prion diseases scrapie in hamster brain homogenates. As these aggregation assays require long-term shaking, BMG LABTECH's microplate readers can be equipped with very robust and durable transport systems.

davin-henderson-thumbnail

Reactive oxygen species assays

Reactive oxygen species (ROS) are endogenously released during immune responses. Exogenously generated ROS can also contribute in diseases due to their DNA-damaging property. A non-fluorescent molecule becoming fluorescent upon oxidation is usually used as a dye for ROS assays. The mechanism, derivatives, and further fluorescent methods for reactive oxygen species detection are outlined in the blog post “Reactive oxygen species detection”. The measurement of ROS is also a useful indicator of mitochondrial toxicity.

 

DNA and RNA quantifications

Fluorescence intensity-based assays that quantify nucleic acids are very popular for sequencing applications. These require exact sample quantification with high specificity, as explained in our blog Next Generation Sequencing (NGS), why quantifying exact DNA/RNA is important. 

RNA and DNA quantification assays are provided for different concentration ranges, with different specificities, with different fluorescent properties, and by many life science chemistry providers. A comparison of different kits and how they perform on BMG LABTECH readers is found in the application note “DNA quantification using absorbance and fluorescent methods."

 

Calcium assays

Calcium ions indicate activation of membrane receptors and regulate muscle contraction and blood clotting. Fluorescence intensity-based assays are employed to study any of these as described in our blog post “Calcium assays: at the centre of biology”. The analysis of intracellular calcium changes requires a fluorescent microplate reader like the CLARIOstar Plus with high temporal resolution as shown in the application note “Real-time calcium flux measurements in iPSC derived 3D heart tissue”, as well as with the possibility to keep cells at 37 °C and 5% CO2.

Additional application could include for instance FRET biosensors to assess ATP levels in living plants experiencing low oxygen conditions.

 

Frequently asked questions

  • What does a microplate reader offer for fluorescence measurements?

    A microplate reader offers rapid data collection, high sensitivity, and multiplexing capabilities for fluorescence measurements. It facilitates quantitative analysis across multiple samples simultaneously.
  • What is fluorescence intensity and why is it important?

    Fluorescence intensity measurement quantifies the emitted light from a fluorescent sample when excited by a specific wavelength. It is crucial for understanding molecular interactions, cellular processes, and biomolecular concentrations in various scientific fields.
  • What types of samples can be analyzed using fluorescence intensity methods?

    A wide range of samples including cells, tissues, biomolecules, and environmental samples can be analyzed to assess their fluorescence properties and quantify specific analytes.
  • What are common applications of fluorescence intensity measurement in research?

    Applications include drug discovery, cell viability assays, gene expression analysis, enzyme activity studies, and environmental monitoring.
  • What factors can affect the accuracy of fluorescence intensity measurements?

    Fluorescence intensity measurements can be influenced by factors including instrument calibration, sample purity, photobleaching effects, ambient light interference, and proper assay protocol adherence.
  • Is it possible to perform kinetic studies using fluorescence intensity measurements?

    Absolutely. Real-time kinetic studies are feasible with suitable instrumentation that allows continuous monitoring of fluorescence changes over time.
  • What units are used for fluorescence intensity?

    Units vary by instrument and software; common outputs are relative fluorescence units (RFUs), counts, or corrected optical units. Absolute quantification needs a standard curve.
  • What is photobleaching and how does it affect measurements?

    Photobleaching is irreversible loss of fluorescence due to light exposure; it reduces intensity over time and should be minimized by limiting exposure and using appropriate settings.
  • What is the difference between fluorescence intensity and fluorescence lifetime?

    Intensity measures emitted light strength; lifetime measures how long a fluorophore stays excited before emitting a photon—two different but complementary readouts.
  • Can fluorescence intensity be used for quantification?

    Yes, with a proper standard curve, validated assay range, and appropriate background corrections.

References

  1. Lakowicz JR, Principles of fluorescence spectroscopy, third edition, 2010
  2. Juan Carlos Stockert, Alfonso Blázquez-Castro (2017). "Chapter 3 Dyes and Fluorochromes". Fluorescence Microscopy in Life Sciences. Bentham Science Publishers. pp. 61–95. ISBN 978-1-68108-519-7.
  3. Bajar BT, Wang ES, Zhang S, Lin MZ, Chu J. A Guide to Fluorescent Protein FRET Pairs. Sensors (Basel). 2016;16(9):1488. Published 2016 Sep 14. doi:10.3390/s16091488
Newsletter Sign-up