Path Length Correction
Path Length Correction in photometers

For many assays such as DNA quantification and protein determination, users frequently compare absorbance results from a microplate reader with those from a 1 cm path length cuvette spectrophotometer.

All absorbance capable microplate readers from BMG LABTECH have a 1 cm path length correction built in via an internal fitting algorithm. Using this correction technique, the 1 cm data from a microplate assay is given with no calculations or estimations made by the user. The path length correction feature can be turned on or off via software as your needs dictata.

A linear regression comparison of data from a BMG LABTECH microplate reader and 1 cm cuvette spectrophotometer produces an r² value of over 0.99 in the UV and visible portions of the spectrum.
Path length correction in plate readers


Jump to top of page
BMG LABTECH GmbH, Hanns-Martin-Schleyer-Str. 10, D-77656 Offenburg/Germany
Phone: +49 781 96968-0, Fax +49 781 96968-67, Impressum